Magna 1 cc

หมวดหมู่:

รายละเอียด

Designed Especially for Mammographic Measurements

The Magna 1cc ionisation chamber is designed especially for mammographic dose and dose rate measurements. It also has an excellent energy response and can be used for radiographic applications. The Magna Ion Chamber is the ideal choice for two reasons; Evaluations of Magna chambers show an average response is within ± 1% over an extended mammography range of 20-40 kVp (HVL of 0.15 mm to 0.73 mm Al).

All mammographic mean glandular dose tables are based on in air measurements. Magna’s air equivalent construction makes it ideally suited for in air exposure measurements of mean glandular dose. The Magna 1 cc will be delivered in a durable case for its best protection.

 

Art. No 9706100-00

Ranges with Piranha and Cobia Flex
(Chamber Adapter required)

Dose Rate 0.25 mGy/s to 2.5 Gy/s
Inaccuracy ±6 % or ±0.025 mGy/s
Energy dependence Within ±1 % for HVL of 0.15 mm to 0,73 mm Al
Within ±2 % for 20 to 150 kVp
Construction Air equivalent conductive plastic
Sensitivity 38 nC/Gy or 0.33 nC/R
Collecting Volume  1 cm3
Ventage Yes
Leakage <50 fA
Typical Bias 300 V DC
Window Kapton conductive film + mylar protective film
Cable length 2 m
Connector LEMO